Total arthroplasty of basal thumb joint with Elektra prothesis: an in vitro analysis

, ,

Journal of Hand Surgery (European Volume): Journal of the British Society for Surgery of the Hand & Official Journal of the Federation of European Societies for Surgery of the Hand

Published online on

Abstract

The reported outcomes of the Elektra thumb carpo-metacarpal joint implant have been very variable. This study evaluates the influence of daily cyclic loads and the type of the screw-fit cup insertion technique in the trapezium, with and without prior threading, on the structural bone behaviour. The study was performed experimentally to predict initial implant stability and cortical bone strains. Computational models were developed to assess the structural cancellous bone behaviour. The use of Elektra implant considerably changed the bone strain behaviour compared with the intact joint. This may be associated with risks of cancellous bone fatigue failure due to overload, particularly in the trapezium. The joint load magnitude has a more important structural role than that of the screw-fit cup insertion technique. Limiting the magnitude of thumb loads after arthroplasty may contribute positively to the longevity of this procedure.

Level of evidence: V