Comparing smartphone camera adapters in imaging post-operative cataract patients
Journal of Telemedicine and Telecare
Published online on February 04, 2016
Abstract
The goal of this study is to compare image quality and clinical confidence for managing post-operative cataract patients based on anterior segment smartphone images obtained in real-world settings using four types of adapters: (a) macro lens (ML), (b) ML with augmented light-emitting diode (LED) illumination (ML-LED), (c) no adapter (NA) and (d) slit lamp (SL) adapter.
Anterior segment images were obtained from 190 eyes after cataract surgery using an eight-megapixel iPhone 6 smartphone camera with four adapters: ML, ML-LED, NA, and SL. Smartphone images were subjectively rated by ophthalmologists as acceptable or not acceptable for: (a) image quality for evaluating the anterior segment structures and (b) reader confidence in clinically managing post-operative patients based on smartphone images.
NA, ML-LED, and SL had the highest scores for image quality with 100%, 93.7%, and 86.3% judged as acceptable, respectively. NA, SL, and ML-LED were also rated highest in clinical confidence with 100%, 98%, and 93.2% having acceptable levels, respectively. ML was judged lowest in both image quality (61.1% acceptable) and clinical confidence (37.4% acceptable).
This study represents the first effort to compare different smartphone camera adapters’ ability to image the anterior segment of the eye in a real-world setting. Our study shows that ML-LED, NA, and SL adapters were acceptable for visualizing anterior segment structures to physician readers in 86–100% of cases. When coupled with visual acuity, intro-ocular pressure and history, these images can result in acceptable clinical confidence in 93–100% of cases.