MetaTOC stay on top of your field, easily

Ten year follow-up of pyrocarbon implants for proximal interphalangeal joint replacement

, , , ,

Journal of Hand Surgery (European Volume): Journal of the British Society for Surgery of the Hand & Official Journal of the Federation of European Societies for Surgery of the Hand

Published online on

Abstract

Results of anatomical resurfacing of the proximal interphalangeal joint using pyrocarbon implants showed reasonable clinical results with a high radiographic migration rate. The aim was to investigate the subjective, clinical, and radiographic results 10 years following surgery, and to compare them with our 2-year follow-up data. We re-evaluated 12 patients with 15 proximal interphalangeal implants on average 9.7 years after surgery. Pain significantly improved from 7.6 on a visual analogue scale pre-operatively to 1.4 at 2 years, and to 0.7 at the final follow-up. The mean total range of motion in all replaced joints was 36° pre-operatively and 39° at the 2-year follow-up, but had decreased significantly to 29° at 10 years. We saw one implant migration in addition to the eight migrated implants we already found 2 years after surgery. The moderate clinical results, combined with the high migration rate, mean that we no longer use this kind of implant.